Draeger Tube, Fluorine 0.1/A 0.1-2ppm, Box of 10
DRAEGER SAFETY INC.
/ Model #8101491
Details
Draeger Tube - Fluorine 0.1/a
Order No. 81 01 491
Application Range | |
Standard Measuring Range | 0.1 to 2 ppm |
Number of Strokes n: | 20 |
Time for Measurement: | approx. 5 minutes |
Standard Deviation: | ± 15 to 20% |
Color Change: | white to yellow |
Ambient Operating Conditions | |
Temperature: | 10 to 40° C ~ 50 to 104° F |
Absolute Humidity: | < 10 mg H2O / L |
Reaction Principle | |
a) F2 + MgCI2 -> CI2 + Mg F2 b) CI2 + o-tolidine yellow reaction product |
|
Cross Sensitivity | |
Nitrogen dioxide, chlorine and chlorine dioxide are indicated, but with different sensitivities. | |
Extension of the Measuring Range | |
Using n = 40, divide the reading by 2; the measuring range will be 0.05 to 1 ppm. |