Draeger Tube, Acetone 40/a, 40-800 ppm, Box of 5
DRAEGER SAFETY INC.
/ Model #8103381
Details
Draeger Tube - Acetone 40/a
Order No. 81 03 381
Application Range | |
Standard Measuring Range | 40 to 800 ppm |
Number of Strokes n: | 1 |
Time for Measurement: | Approximately 1 minute |
Standard Deviation: | ± 15 to 20% |
Color Change: | pale yellow → yellow |
Ambient Operating Conditions | |
Temperature: | 5 to 40° C ~ 41 to 104° F |
Absolute Humidity: | 5 to 40 mg H2O / L |
Reaction Principle | |
Acetone + 2.4-Dinitrophenylhydrazine → yellow hydrazone | |
Cross Sensitivity | |
Other ketones are indicated but with different sensitivities. Aldehydes are indicated. 500 ppm ethylacetate does not affect the indication. Ammonia causes the indicating layer to turn yellowish brown. |