Draeger Tube, Olefins .05%/A 0.04-3.2 Vol%, Box 10

DRAEGER SAFETY INC. / Model #CH31201
draeger tube olefins
Clearance
$118.00
Availability: May direct ship from manufacturer.
Share:
Details

Draeger Tube - Olefine 0.05%/a
Order No. CH 31 201

Application Range
Standard Measuring Range 0.06 to 3.2 vol. % propylene
0.04 to 2.4 vol. % butylene
Number of Strokes n: 20 to 1
Time for Measurement: max. 5 minutes
Standard Deviation: ± 30%
Color Change: violet to pale brown
Ambient Operating Conditions
Temperature: 0 to 40° C ~ 32 to 104° F
Absolute Humidity: < 30 mg H2O / L
Reaction Principle
CH3-CH2-CH=CH2 + MnO4- → MnIV + various oxidation products
Cross Sensitivity
Many organic compounds with C=C double bonds are indicated, but with different sensitivities. It is impossible to differentiate between them. It is impossible to measure olefines in the presence of dialkyl sulfides.


View the entire Draeger Tubes & CMS Handbookk

Recently Viewed

Draeger Tube, Olefins .05%/A  0.04-3.2 Vol%, Box 10
DRAEGER SAFETY INC.
Model #CH31201
Draeger Tube, Olefins .05%/A 0.04-3.2 Vol%, Box 10
$118.00