DraegerTube, Epichlorihydrin 5/C 5-80ppm
DRAEGER SAFETY INC.
/ Model #6728111
Details
Draeger Tube - Epichlorohydrine 5/c
Order No. 67 28 111
Application Range | |
Standard Measuring Range | 5 to 80 ppm |
Number of Strokes n: | 20 |
Time for Measurement: | approx. 8 minutes |
Standard Deviation: | ± 15 to 20% |
Color Change: | pale gray to yellow orange |
Ambient Operating Conditions | |
Temperature: | 10 to 40° C ~ 50 to 104° F |
Absolute Humidity: | 5 to 15 mg H2O / L |
Reaction Principle | |
Epichlorohydrine + CrVI CI2 CI2 + o-tolidine yellow orange reaction product |
|
Cross Sensitivity | |
Other chlorinated hydrocarbons are indicated as well but with different sensitivities. It is not possible to measure epichlorohydrine in the presence of free halogens and hydrogen halides in the TLV range because they are indicated as well. Petroleum hydrocarbons cause low readings. |