Draeger Tube - Styrene 50/a
Order No. CH 27 601
Application Range | |
Standard Measuring Range | 50 to 400 ppm |
Number of Strokes n: | 2 to 11 |
Time for Measurement: | max. 2 minutes |
Standard Deviation: | ± 30% |
Color Change: | white to yellow |
Ambient Operating Conditions | |
Temperature: | 10 to 40° C ~ 50 to 104° F |
Absolute Humidity: | < 15 mg H2O / L |
Reaction Principle | |
C6H5-CH=CH2 + H2SO4 → yellow reaction product | |
Cross Sensitivity | |
Other organic compounds which tend toward polymerization (e. g. butadiene) are indicated. It is impossible to measure monostyrene in the presence of these compounds. |