Draeger Tube, Box of 9 Nickel Tetracarbonyl
DRAEGER SAFETY INC.
/ Model #CH19501

Details
Draeger Tube - Nickel Tetracarbonyl 0.1/a
Order No. CH 19 501
Application Range | |
Standard Measuring Range | 0.1 to 1 ppm Discoloration compared to color standard |
Number of Strokes n: | 20 |
Time for Measurement: | approx. 5 minutes |
Standard Deviation: | ± 50% |
Color Change: | yellow to pink |
Ambient Operating Conditions | |
Temperature: | 0 to 30° C ~ 32 to 86° F |
Absolute Humidity: | < 30 mg H2O / L |
Reaction Principle | |
a) Ni(CO)4 + I2 Nil2 + 4 CO b) Nil2 + Dimethylglyoxime pink colored complex |
|
Cross Sensitivity | |
Iron pentacarbonyl is also indicated by a brownish discoloration, however, with less sensitivity. Measurement of nickel tetra carbonyl is not possible in the presence of hydrogen sulfide of sulfur dioxide, since the reading is suppressed. Such a disturbance can be recognised by decoloration of the indicating layer even before the reagent ampoule is opened. | |
Additional Information | |
After performing the required 20 pump strokes the reagent ampoule must be broken and the liquid carefully drawn onto the indicating layer using the pump. |
Recently Viewed