Draeger Tube - Carbon Monoxide 8/a
Order No. CH 19 701
Application Range | |
Standard Measuring Range | 8 to 150 ppm |
Number of Strokes n: | 10 |
Time for Measurement: | approx. 2 minutes |
Standard Deviation: | ± 10 to 15% |
Color Change: | white to pale brown |
Ambient Operating Conditions | |
Temperature: | 0 to 50° C ~ 32 to 122° F |
Absolute Humidity: | < 50 mg H2O / L |
Reaction Principle | |
5 CO + I2O5 I + 5 CO2 | |
Cross Sensitivity | |
Acetylene is also indicated, however, with less sensitivity. Petroleum hydrocarbons, benzene, halogenated hydrocarbons and hydrogen sulfide are retained in the pre-layer. In the case of higher concentrations of interfering hydrocarbons, use should be made of a carbon pre-tube (CH 24 101). Higher concentrations of easily cleavable halogenated hydrocarbons (e.g. trichloroethylene), are liable to form chromyl chloride in the pre-layer which changes the indicating layer to a yellowish-brown. CO determination is impossible in the case of high olefin concentrations. |