Register to Win the Flying Cross & Vertx Giveaway Package

Draeger Tube, Chloroformates 0.2/B 0.2-10ppm, Box of 10

Draeger Tube, Chloroformates 0.2/B 0.2-10ppm, Box of 10

$123.00
Availability: May direct ship from manufacturer.
Product Code: 6718601

Draeger Tube - Chloroformate 0.2/b
Order No. 67 18 601

Application Range
Standard Measuring Range 0.2 to 10 ppm
Number of Strokes n: 20
Time for Measurement: Approximately 3 min
Standard Deviation: ± 20 to 30%
Color Change: white to yellow
Ambient Operating Conditions
Temperature: 10 to 40° C ~ 50 to 104° F
Absolute Humidity: 5 to 15 mg H2O / L
Reaction Principle
CICOOR + 4-(4-nitrobenzyl)-pyridine  yellow reaction product
Cross Sensitivity
Methyl, ethyl and isopropylchloroformate are indicated with approximately the same sensitivity. It is impossible to differentiate them. Petroleum hydrocarbons, aromatics, alcohols and ketones in the TLV range do not affect the indication. It is impossible to measure chloroformate in the presence of phosgene.


View the entire Draeger Tubes & CMS Handbook

Share: