Draeger Tube - Phosphine 50/a
Order No. CH 21 201

Application Range
Standard Measuring Range 50 to 1,000 ppm
Number of Strokes n: 3
Time for Measurement: approx. 2 minutes
Standard Deviation: ± 10 to 15%
Color Change: yellow to brown black
Ambient Operating Conditions
Temperature: 0 to 50° C ~ 32 to 122° F
Absolute Humidity: < 40 mg H2O / L
Reaction Principle
PH3 + Au3+ → Au (colloidal)
Cross Sensitivity
Arsine and antimony hydride are indicated, but with different sensitivities.

Hydrogen sulfide, mercaptans, ammonia, carbon monoxide, sulfur dioxide and hydrochloric acid in the TLV range do not interfere.
Extension of the Measuring Range
Using n = 10, multiply the reading by 0.3; the range of measurement is 15 to 300 ppm.