Draeger Tube, Sulfur Dioxde 20/A 20-2000ppm, Box of 10
DRAEGER SAFETY INC.
/ Model #CH24201
Details
Draeger Tube - Sulfur Dioxide 20/a
Order No. CH 24 201
Application Range | |
Standard Measuring Range | 20 to 200 ppm |
Number of Strokes n: | 10 |
Time for Measurement: | approx. 3 minutes |
Standard Deviation: | ± 10 to 15% |
Color Change: | brown yellow to white |
Ambient Operating Conditions | |
Temperature: | 0 to 40° C ~ 32 to 104° F |
Absolute Humidity: | < 30 mg H2O /L |
Reaction Principle | |
SO2 + I2 + 2 H2O → H2SO4 + 2 HI | |
Cross Sensitivity | |
Hydrogen sulfide is indicated, with the same sensitivity. It is impossible to measure sulfur dioxide in the presence of hydrogen sulfide. Nitrogen dioxide will shorten the reading.. | |
Extenstion of the Measuring Range | |
Using n = 1+3 desorption strokes, multiply the reading by 10; the range of measurement is 200 - 2,000 ppm. The desorption strokes are to be taken in clean air (i.e. free of sulfur dioxide) immediately following the single pump stroke. |