Draeger Tube - Mercaptan 20/a
Order No. 81 01 871

Application Range
Standard Measuring Range 20 to 100 ppm
Number of Strokes n: 10
Time for Measurement: approx. 2.5 minutes
Standard Deviation: ± 10 to 15%
Color Change: white to yellow brown
Ambient Operating Conditions
Temperature: 0 to 50° C ~ 32 to 122° F
Absolute Humidity: 3 to 30 mg H2O / L
Reaction Principle
a) 2 R-SH + Cu2+ → Cu(RS)2 + 2 H+
b) Cu(RS)2 + S → yellow brown copper compound
Cross Sensitivity
Higher molecular weight alkyl mercaptans (e.g. propyl- and butylmercaptans) are indicated with approximately the same sensitivity.

Hydrogen sulfide is indicated with approximately twice the sensitivity of the mercaptans (e.g. 10 ppm hydrogen sulfide gives an indication of 20 ppm). In presence of Hydrogen Sulfide a measurement of mercaptans is impossible.
Additional Information
After performing the required ten pump strokes the reagent ampoule must be broken. The liquid of the ampoule must be transferred to the indicating layer and carefully drawn through it using the pump. After completing the measurement wait for 3 min prior to evaluation.